DVM645BI GB 21
AC Current
Range resolution Accuracy (% reading + digits)
4mA 1µA ± (1.5% rdg + 5d) 45 to 400Hz
40mA 10µA ± (1.5% rdg + 5d) 45 to 400Hz
400mA 100µA ± (1.5% rdg + 5d) 45 to 400Hz
10A
(20A for 30
seconds)
10mA ± (2% rdg + 10d) 45 to 400Hz
Input protection: 1A/250V fuse for mA input, 15A / 250V fuse for A input
Load voltage: 600mV max. for mA input, 900mV max. for A input
AC coupled true RMS
Resistance
Range resolution Accuracy (% reading + digits)
400Ω 0.1Ω
± (0.5% rdg + 5d)
4kΩ 1Ω
± (0.5% rdg + 3d)
40k
Ω
10
Ω
± (0.5% rdg + 3d)
400k
Ω
100
Ω
± (0.5% rdg + 3d)
4MΩ 1kΩ
± (1% rdg + 5d)
40MΩ 10kΩ
± (1.5% rdg + 10d)
Open circuit Voltage: 0.45V
Input protection: 250V RMS
Continuity Test
Continuity threshold: approx. 40Ω
Continuity threshold: 2kHz buzzer tone
Input protection: 250V RMS
Diode Test
Test current: 0.6mA
Open circuit voltage: approx. 3.0V
Input protection: 250V RMS