2. Specifications
2-1. General Specifications
Display 3 3/4 digit LCD with a max. reading of 4000.
Range Control Manual range control.
Polarity Automatic negative polarity indication.
Zero Adjustment Automatic.
Overrange Indication “OL on LCD is displayed.
Low Battery Indication “ ” sign on LCD readout.
Auto Power Off Approximately 15 minutes.
Safety Standards EMC/LVD. The meter is up to the
standards of IEC1010 Pollution Degree 2,
Overvoltage Catagory II.
Operation Temperature 32
O
F to 104
O
F (0
O
C to 40
O
C)
less than 85% relative humidity
Storage Temperature –4
O
F to 140
O
F (–20
O
C to 60
O
C)
less than 95% relative humidity
Power 9V battery.
Dimensions 7 5/8” (H) x 3 9/16” (W) x 1 3/4” (D)
(w/o holster)
Weight Approximately 0.66 lb. (w/o holster)
Accessories Safety Test Leads - 1 pair
Operator’s Manual - 1 pc.
Holster - 1 pc.
2-2. Electrical Specifications
Accuracy is + (% or reading + number in last digit) at 23 + 5
O
C, <75% RH.
DC Voltage
400mV, 4V, 40V, 400V +(0.5% + 3)
1000V +
(0.8% + 3)
Impedance 10MW
Max. input: 250V~ on 400mV range; 1000V~ on all other ranges.
AC Voltage
4V, 40V, 400V +(0.8% + 3)
400mV, 750V +(1.2% + 3)
Impedance 10MW
Max. input: 250V~ on 400mV range; 1000V~ on all other ranges.
Resistance
400W, 4kW, 40kW, 400kW, 4MW +
(1% + 3)
40MW +
(2% + 3)
400MW +
(5% + 10)
Overload Protection 250V DC/ACrms
DC Current
40mA, 400mA, 4000mA+
(1.2% + 3)
40mA, 400mA +
(1.5% + 3)
10A +(2% + 5)
Caution: Max. operation time 15s.
Overload Protection Fast 0.5A/250V, 10A/250V fuse.
AC Current
40mA, 400mA, 4000mA+
(1.5% + 3)
40mA, 400mA +
(2% + 3)
10A +(2.5% + 5)
Caution: Max. operation time 15s.
Overload Protection Fast 0.5A/250V, 10A/250V fuse.
Frequency Response 40 - 400Hz
Capacitance
4nF +
(5% + 10)
40nF +(3% + 10)
400nF, 4mF, 40mF+(2% + 5)
200mF+
(4% + 5)
Overload Protection 250V DC/ACrms
Frequency
Frequency 10Hz - 10MHz +(0.1% + 5)
Sensitivity Sine wave 0.6Vrms
Overload Protection 250V DC/ACrms
Duty Cycle
Duty Cycle 0.1% - 99.9% +
(2% + 2) Frequency lower than 10kHz
Sensitivity Sine wave 0.6Vrms
Overload Protection 250V DC/ACrms
Diode Test
Test Current: 1 +
0.6mA Test Voltage: Approx. 1.5V
Overload Protection 250V DC/ACrms
Continuity Test
Audible Indication Less than 120W approx.
Overload Protection 250V DC/ACrms
h
FE Test
Ib 10mA
Vce 2.5V approx.
Test Range 0 - 1000
Logic Test
TTL Logic 1: 2.5V +
0.8V Logic 0: 0.8V +0.5V
CMOS Logic 1: 4.0V +1.0V Logic 0: 2.0V +0.5V
Frequency Response: 20MHz
Detectable Pulse Width: 25ns min
-4- -5-
+
Max. voltage from “VWmA”
socket to “COM” socket is
1000V~ from “COM” socket
to earth is 500V~.
!
!